Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
►
Unplug the device before cleaning.
DANGER
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions. The cleaning
and disinfecting agent must be approved for
plastics and high-grade steel, and it must not
attack softeners (e.g. silicone).
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum permitted cleaning
temperature of 100 °C.
Note
There may be significant discoloration of the camera head when using
Cidex OPA. However, this discoloration has no influence on the functional-
ity or safety of the product.
6.6
Manual cleaning and disinfecting
Post use and pre-cleaning activities
►
Pre-clean the device immediately after use.
►
Remove and discard the sterile drape, if present.
►
Decouple the endoscope from the optocoupler.
►
Uncouple the camera head from the controller.
►
Pre-clean all used parts of the fully disassembled product at the oper-
ating table at the conclusion of the surgery by wiping with a lint-free
wipe wetted (i.e. wetted but not dripping; squeeze out excess solution)
with an enzymatic cleaning solution until visibly clean. Before wetting
the lint-free wipe with the enzymatic clean solution, the solution
should be prepared according to the manufacturer's instruction.
►
Arrange for reprocessing, ensuring that all of the device components
are reprocessed within 6 hours.
12
®