2D camera heads PV481, PV482, PV485
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the device before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Use cleaning and disinfecting agents according
to the manufacturer's instructions. The cleaning
and disinfecting agent must be approved for
plastics and high-grade steel, and it must not
attack softeners (e.g. silicone).
Observe specifications regarding concentration,
temperature and exposure time.
Do not exceed the maximum permitted cleaning
temperature of 100 °C.
Note
There may be significant discoloration of the camera head when using
Cidex OPA. However, this discoloration has no influence on the functional-
ity or safety of the product.
6.6

Manual cleaning and disinfecting

Post use and pre-cleaning activities
Pre-clean the device immediately after use.
Remove and discard the sterile drape, if present.
Decouple the endoscope from the optocoupler.
Uncouple the camera head from the controller.
Pre-clean all used parts of the fully disassembled product at the oper-
ating table at the conclusion of the surgery by wiping with a lint-free
wipe wetted (i.e. wetted but not dripping; squeeze out excess solution)
with an enzymatic cleaning solution until visibly clean. Before wetting
the lint-free wipe with the enzymatic clean solution, the solution
should be prepared according to the manufacturer's instruction.
Arrange for reprocessing, ensuring that all of the device components
are reprocessed within 6 hours.