Mechanical alkaline cleaning and thermal disinfecting
Phase
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection 90/194
V
Drying
Use a washer/disinfector that has been validated as effective and
meets the performance requirements of ISO 15883-1 (or the respective
country specific version thereof)
For reprocessing, use the camera tray, which is used without a lid. This
tray is only suitable for the camera and not for instruments.
Place the device in the provided reusable camera tray. Avoid areas that
cannot be reached during rinsing.
Place the tray in the automatic washer/disinfector in a suitable carrier
that ensures both the protection of the instrument and access to sur-
faces for cleaning. No lid will be used for the tray.
Start the cleaning cycle as described (see table) in accordance with the
manufacturer's instructions and instruction for use for the washer/dis-
infector.
Remove the tray and the devices from automatic washer.
Visually inspect the device in a well-lit area; they should be completely
clean, dry and undamaged.
Use a magnifying glass. Repeat the cleaning process if necessary.
Immediately set aside any damaged components.
6.8

STERRAD sterilisation

The STERRAD® sterilization systems, manufactured by Advanced
Sterilization Products (ASP), use low-temperature, hydrogen peroxide gas
plasma technology for terminal sterilization of properly cleaned, rinsed,
and dried reusable medical devices.
Refer to the STERRAD® Sterilization Systems User Guide for detailed
instructions for use of any STERRAD® unit and ASP's STERRAD Sterility
Guide (SSG) at www.sterradsterilityguide.com or contact ASP cus-
tomer service.
T [°C/°F]
t [min]
<25/77
3
55/131
10
>10/50
1
5
Water quality
Chemicals/Note
Tap water
Deionized water Alkaline Detergent
Neodisher Mediclean forte 0.5 %
Deionized water –
Deionized water –
According to the program of the washer/disinfector
Performing sterilization
Note
STERRAD® sterilization may result in cosmetic device changes that do not
necessarily affect the device function.
Loads containing moisture may cause a cycle can-
cellation!
CAUTION
Place all parts in a cleaning and storage basket.
Place a STERRAD® indicator strip in the basket.
Assemble sterilization container JM441 according to manufacturer's
instruction.
Place the basket in the sterilization container JM441.
Load the container in the sterilizer:
– Position the container such that the plasma can fully surround it.
– Load only one container per cycle in the chamber.
– Leave the other shelf empty.
Start the sterilization cycle in accordance with the manufacturer's
instructions and instructions for use for the steam sterilizer.
STERRAD® sterilization was validated for the following cycles:
– STERRAD® 100S short cycle
– STERRAD® NX standard cycle
– STERRAD® 100NX standard cycle
Remove the sterilized products from the sterilizer.
Ensure that the products remain sterile after reprocessing.
This concludes sterilization.
Thoroughly dry the product before loading into
the STERRAD® sterilizer.