STERILIZATION
See Table 2 for typical sterilization parameters that have been qualified for the sterilization of Sonata IUUS Probe Reprocessing Tray.
NOTE: These sterilization methods and cycles are validated for the tray ONLY. Refer to the Sonata® IUUS Probe
Instructions for Use (REF-003, IUSP-002 Manual) for validated sterilization methods for the IUSP-002 IUUS Probe
in the tray OM-1000-GS.
Table 2.
STERILIZATION METHOD FOR TRAY
STERRAD 100S
STERRAD 100NX
STERRAD 100NX
STERRAD 100NX
STERRAD NX
STERIS AMSCO V-PRO 1
STERIS AMSCO V-PRO 1 PLUS
STERIS AMSCO V-PRO maX
STERIS AMSCO V-PRO maX 2
Summit Medical has validated the following sterilization methods:
The STERRAD 100S Standard sterilization cycle in legally marketed wrap cleared by the FDA.
The STERRAD 100NX Standard sterilization cycle in legally marketed wrap cleared by the FDA.
The STERRAD 100NX Express sterilization cycle in legally marketed wrap cleared by the FDA.
The STERRAD 100NX Flex sterilization cycle in legally marketed wrap cleared by the FDA.
The STERRAD NX Standard sterilization cycle in legally marketed wrap cleared by the FDA.
The STERIS AMSCO V-PRO, V-PRO 1 PLUS, V-PRO maX and V-PRO maX 2 sterilization cycles in legally marketed wrap cleared by the FDA.
Use an FDA cleared accessory to maintain sterility.
Please consult the sterilizer instruction manual to ensure intended loads are compatible with the intended sterilization cycle.
Ensure IUUS Probe Reprocessing Tray and enclosed instrument are completely dry before STERRAD or STERIS V-PRO sterilization.
CYCLE
(times)
Standard
Standard
Express
Flex
Standard
Standard
Lumen, Non-Lumen
Lumen, Non-Lumen
Lumen, Non-Lumen
3