Ambu aScope Duodeno Manual Del Usuario página 16

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 160
Connection to Fluid Canisters 6
To perform patient examinations or procedures, all fluid containers (e.g. sterile water bottle)
must be properly and securely connected to the aBox™ Duodeno to prevent spillage, thereby
maintaining a safe work environment. Place the containers in the designated locations and
connect them according to the instructions in this chapter. Please note, that a new suction
canister/sterile water bottle must be used for each procedure.
CO₂ Connection/ Connection to the Insufflation System 6
The aScope™ Duodeno is designed to work with a source of medical grade CO₂ for insufflation.
The maximum medical grade CO₂ supply pressure is 12 psi (80 kPa). Connect to the aScope™
Duodeno using a tube with a luer-lock connector.
If ancillary equipment is ON, turn OFF.
Connect the luer-lock of the insufflation tubing from the insufflator to the insufflation
connector of the aScope™ Duodeno.
Turn the ancillary equipment back ON.
Connection to the Suction System
Regardless of the vacuum source chosen, the aScope™ Duodeno will require a source to
provide a vacuum of min. -7 psi (-50 kPa) and max. -11 psi (-76 kPa) for the endoscope to operate
normally. Failure to provide the minimum vacuum requirements could result in a decreased
capacity in removing patient waste and/or irrigation fluid and loss of the cooling function.
If ancillary equipment is ON, turn OFF.
Fit the end of the suction tubing securely over the suction connector located on the
endoscope connector of the aScope™ Duodeno.
Connect the other end of the suction tubing to the external suction system (e.g. wall
suction or medical suction pump).
Turn the ancillary equipment back ON.
Electrical Connection to the aBox™ Duodeno
Make sure all components as described above have been connected, the ancillary devices
and the aBox™ Duodeno has been set and powered up.
The aBox™ Duodeno indicates when it is ready by displaying the information message on
the main monitor screen: "please connect endoscope". For detailed information about
setting-up the aBox™ Duodeno, please refer to the IFU of the specific device.
Lift the endoscope connector cover on the aBox™ Duodeno and plug the endoscope
connector into the connector socket of the aBox™ Duodeno.
The system will ask you then to confirm that a new bottle of sterile water was prepared
for the procedure.
Once confirmed the system will display the live procedure image. For detailed information
about using the aBox™ Duodeno, please refer to the IFU of the specific device.
The illumination must be turned ON by the user. For detailed information about using the
aBox™ Duodeno, please refer to the IFU of the aBox™ Duodeno.
Inspection of the Image
Turn on the LED by pressing the illumination button on the aBox™ Duodeno.
Verify that a live video image appears on the screen by pointing the distal end of the
aScope™ Duodeno towards an object, e.g. the palm of your hand and make sure that the
orientation of the image is correct.
Adjust the image preferences, if necessary, by accessing the settings and information
menu of the aBox™ Duodeno. For detailed information, please refer to the aBox™
Duodeno IFU.
Angulate the endoscope distal end and confirm there are no disruptions to the live
image, nor are there any other irregularities or noted artifacts.
If the image is impaired and/or unclear, wipe the lens at the distal end using a sterile cloth.
16
6
8
7

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido