* On some monitors, S-VHS inputs may be labeled Y/C.
** The DVI connectors can also output analog SXGA signals through a DVI-I to VGA adapter.
When using any device with unterminated analog video inputs, connect a cable from the VIDEO OUT of that device to the
VIDEO IN on the monitor.
3. Connect the DVI output from the Stryker digital capture console to the DVI input on the display
monitor. (The VisionPro 26" LED Display (0240-031-020) is shown as an example in the
following diagram.)
4. Using the provided remote cables, connect remote outputs 1 and 2 from the 1588 AIM console
to a Stryker digital capture console. (The remote inputs should be connected to the same
channel as the video input.)
•
Devices connected to the remote outputs can be operated using the console touchscreen
or the P button on the camera head. See the Operation section for details.
•
Once connected to the 1588 AIM Camera, the SDC3 can control ENV, Dynamic Range
Enhancement, and Strobe functions. The user can also customize button configurations
through the SDC3.
5. Connect a USB A-to-A cable from the Light Source output on the 1588 AIM console to the
CCU input on the L10 LED Light Source with AIM Technology (0220-220-300) or the Precision
LED Light Source (0220-220-000). It is recommended to use the USB cable (P30883)
provided with the Stryker light source, as use of third-party cables may prevent the devices
from properly communicating.
•
To use ENV mode, the 1588 AIM Camera requires a connection to the L10 LED Light
Source with AIM Technology.
•
The 1588 AIM Camera Head can be programmed to toggle Run/Standby controls on the
light source. Contact a Stryker representative for more information about enabling this
advanced feature.
EN
12
P29924F