P P r r o o b b l l e e m m
SIDNE device does not
recognize camera head
ENV mode not enabled
N N o o t t e e - - If this Troubleshooting section does not resolve the problem, call Stryker Technical
Support at 1-877-478-7953 (inside the U.S.) or refer to the standard warranty.
EN
30
P P o o s s s s i i b b l l e e S S o o l l u u t t i i o o n n
Contact your Stryker representative for compatibility
settings.
Confirm the camera console is set to the
Laparoscopy Surgical Specialty.
Confirm a USB A-to-A cable is connected from the
camera console to the L10 LED Light Source with
AIM Technology
Confirm white light is activated on the light source
P29924F