4. The camera head surface may exceed 41 °C (106 °F) in operating conditions with high
ambient temperatures and it should be handled with caution.
5. The camera head and coupler are shipped non-sterile. You must sterilize these devices before
the first use and after each use. To prevent device damage and infection risk to the patient or
user, follow all cleaning and sterilization instructions in this manual.
6. Do not position the console so that it is difficult to disconnect the power cord from the supply
mains.
7. To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
8. Multiple portable socket-outlets shall not be placed on the floor.
9. Never use the camera system in the presence of flammable or explosive gases.
10. Disconnect the console from the electrical outlet when inspecting fuses.
11. Do not remove covers on the console, as doing so may cause damage to electronics and/or
electric shock.
12. Do not disassemble any part of the camera head; doing so may break the seals, causing
leakage and/or electric shock.
13. Attempt no internal repairs or adjustments not specifically detailed in this operating manual.
W W a a r r n n i i n n g g s s : : E E N N V V M M o o d d e e
W W A A R R N N I I N N G G - - I I M M P P O O R R T T A A N N T T S S A A F F E E T T Y Y N N O O T T I I C C E E - - L L A A S S E E R R R R A A D D I I A A T T I I O O N N : :
•
Endoscopic Near-Infrared Visualization (ENV) mode controls a Class 1M laser emitted from
the L10 LED Light Source with AIM Technology (0220-220-300).
•
Use of controls or performance of procedures other than those specified herein can result in
hazardous laser radiation exposure and can cause severe eye injury to the patient or user.
•
To avoid exposure to laser radiation, follow all warnings and guidelines presented below and
throughout this user manual.
1. Before using ENV mode, read and be familiar with all instructions and warnings found in this
user manual and in user manual P27006 (English) or P27009 (multilingual) for the L10 LED
Light Source with AIM Technology.
2. Protect the 1588 AIM Video Camera against unqualified use.
3. Wear eye protection as appropriate. Refer to any applicable regional regulations or standards
for personal protective equipment.
4. Do not manipulate tissue while ENV mode is enabled. The camera is intended to visualize
tissue manipulation only while ENV mode is off.
5. When using ENV mode, do not view the light output with optical instruments (for example,
microscopes or magnifiers). Do not direct the light output in ENV mode into an area where
such instruments are likely to be used.
6. Do not enable ENV mode when the endoscope is outside of the patient's body.
7. When ENV mode is enabled, never look into the following apertures or direct the light emitted
from the apertures toward another person:
•
the light cable connection on the light source (if the cable is not attached)
•
the end of the light cable (if the SafeLight adapter is attached)
EN
4
P29924F