3. Push in the connector until it locks in place.
N N o o t t e e - - To unplug the camera from the console, grasp the knobbed portion of the connector and
pull straight out.
S S e e t t t t i i n n g g U U p p t t h h e e C C o o u u p p l l e e r r
Steps 1–3 below provide instructions for connecting 1588 AIM Camera Heads to the AIM Coupler,
18 mm, C-Mount (1588-020-122). Refer to the bullets below for possible system variations:
When using the 1588 AIM Camera Head with Integrated Coupler (1588-610-122), skip to step
2.
When using the 1588 Pendulum Camera Head with Integrated Coupler (1588-310-130), see
Stryker user manual P29925 (English) or P29926 (multilingual).
When using a direct-coupled C-Mount endoscope (a scope that requires no coupler), thread
the endoscope directly into the camera head until it forms a tight seal, and skip to step 3. (C-
Mount endoscopes are not compatible with camera heads that have an integrated coupler.)
W W A A R R N N I I N N G G - -
When attaching or removing the coupler, grip only the rear adapter, as twisting other parts of
the coupler with force may result in mechanical damage.
Do not overtighten the coupler (or a direct-coupled C-mount scope), as this may damage the
front window of the camera.
1. Attach the coupler to the camera head.
Gripping the rear adapter, screw the coupler clockwise onto the camera head until it forms
a tight seal (1588-210-105 and 1588-710-105 only).
EN
14
P29924F