Flush any crevices and mated surfaces 5 times. After all visible detergent residue is
removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it with a clean cloth or filtered pressurized air.
Visually inspect the device for cleanliness, paying close attention to hard-to-reach areas. If
visible soil remains, repeat steps 1– 3.
4. S S o o a a k k
Prepare a non-enzymatic detergent according to the manufacturer instructions.
Fully immerse the device in the detergent. Use a syringe to inject 50 mL of the detergent
into any crevices and mated surfaces.
Soak the device in the detergent for 15 minutes.
5. B B r r u u s s h h
Thoroughly brush the exterior of the device with a soft-bristle brush for 90 seconds,
focusing on any mated or rough surfaces.
Use a syringe to inject 50 mL of the detergent into any crevices and mated surfaces
5 times.
When cleaning the 1588 AIM Camera Head with Integrated Coupler (1588-610-122), brush
between all gaps and crevices while pushing down on the endobody clamp. Continue
brushing all gaps and crevices while releasing the clamp to the initial position.
6. R R i i n n s s e e
Remove the device from the prepared detergent. Rinse the device with reverse osmosis/
deionized water at ambient temperature for 90 seconds or until all visible detergent residue
is removed.
Flush any crevices and mated surfaces 5 times. After all visible detergent residue is
removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it with a clean cloth or filtered pressurized air.
Visually inspect the device for cleanliness, paying close attention to hard-to-reach areas. If
visible soil remains, repeat steps 1– 6.
A A u u t t o o m m a a t t e e d d C C l l e e a a n n i i n n g g
N N o o t t e e - - For necessary materials and equipment, see the Materials and Equipment table.
1. S S o o a a k k
Disassemble the camera head from the scope and coupler.
Ensure the soaking cap is installed. Refer to the Installing the Soaking Cap section for
more detail about installing the cap.
Prepare an enzymatic detergent according to the manufacturer instructions.
Use a clean cloth to wipe the entire device with the detergent.
Fully immerse the device in the detergent. Use a syringe to inject 50 mL of the detergent
into any crevices and mated surfaces to remove loose debris.
Soak the device in the detergent for 15 minutes.
2. B B r r u u s s h h
Thoroughly brush the exterior of the device with a soft-bristle brush for 90 seconds,
focusing on any mated or rough surfaces.