T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g
P P r r o o b b l l e e m m
E1 error code
E2 error code
E3 error code
E4 error code
Touchscreen freezes
"Restart Camera Console"
message (Color bar
background)
"System Error" message
(Light blue background)
No color bar
Incorrect picture color
P29924F
P P o o s s s s i i b b l l e e S S o o l l u u t t i i o o n n
Turn off the console, wait 3 seconds, and turn it back
on.
Turn off the console, wait 3 seconds, and turn it back
on.
Turn off the console, wait 3 seconds, and turn it back
on.
Ensure vent holes on console are not obstructed.
Turn off the console, wait 3 seconds, and turn it back
on.
Turn off the console, wait 3 seconds, and turn it back
on.
Camera head temporarily shut down due to
overcurrent. Turn off the console, wait 3 seconds,
and turn it back on.
After sterilization, ensure the camera head has
cooled down before connecting it to the console.
No video detected.
After sterilization, ensure the camera head has
cooled down before connecting it to the console.
Return the system for repair.
Ensure the video-out from the console is connected
to the video-in on the monitor.
Ensure all video systems are powered on.
Ensure that the camera head is not connected to the
console.
Turn off the console, wait 3 seconds, and turn it back
on.
Perform the White Balance test. (See the Performing
the White Balance Test section.)
Check the color settings on the monitor.
27
EN