Descargar Imprimir esta página

Stryker 1588 AIM Manual Del Usuario página 40

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 290
5. Thoroughly rinse and flush the device with running, reverse osmosis/deionized water to
remove the disinfectant.
6. Dry the device with a clean, lint-free cloth immediately after rinsing.
S S t t e e r r i i l l i i z z a a t t i i o o n n
After performing the cleaning instructions specified above, perform one of the following
sterilization cycles.
N N o o t t e e - - For necessary materials and equipment, see the Materials and Equipment table.
S S t t e e r r r r a a d d
1. Clean and prepare the device as recommended in this user manual. Ensure the soaking cap is
installed.
2. If using a sterilization tray (optional), follow any additional instructions provided with the tray.
Use only trays that are compatible with Sterrad.
3. Double wrap the device (or tray) prior to sterilization.
4. Sterilize the device following the instructions of the manufacturer, using the Sterrad 100S, 200,
NX, or 100NX Sterilization System. Select the standard cycle.
5. After sterilization, allow the device to cool to room temperature before reconnecting it to a
coupler or the console. Otherwise, the lens can fog during use or the console can produce a
system error.
S S t t e e r r i i s s / / A A m m s s c c o o V V - - P P R R O O
1. Clean and prepare the device as recommended in this user manual. Ensure the soaking cap is
installed.
2. If using a sterilization tray (optional), follow any additional instructions provided with the tray.
Use only trays that are approved for sterilization with V-PRO.
3. Double wrap the device (or tray) prior to sterilization.
4. Sterilize the device using one of the following V-PRO sterilization systems:
V-PRO 1 (Standard cycle)
V-PRO 1 Plus (Non-Lumen or Standard cycle)
V-PRO maX (Non-Lumen or Standard cycle)
5. After sterilization, allow the device to cool to room temperature before reconnecting it to a
coupler or the console. Otherwise, the lens can fog during use or the console can produce a
system error.
S S t t e e r r i i s s S S y y s s t t e e m m 1 1 / / 1 1 E E / / 1 1 P P l l u u s s / / 1 1 E E x x p p r r e e s s s s
N N o o t t e e - - Steris System 1, System 1 Plus, and System 1 Express are not intended for use in the
United States.
1. Clean and prepare the device as recommended in this user manual. Ensure the soaking cap is
installed.
2. Following the instructions of the manufacturer, sterilize the device using one of the Steris
systems below with the appropriate sterilant:
System 1 with Steris 20 Sterilant
EN
38
P29924F

Publicidad

loading

Este manual también es adecuado para:

1588-010-0001588-210-1051588-610-1221588-710-105