P P r r o o b b l l e e m m
White Balance quality is not
good
Picture is too dark
Picture is too bright
Noise or snow on picture
when using electrocautery
probes
Noise or snow on picture
when not using
electrocautery probes
No video picture when the
camera head is plugged in
Image is not well centered
EN
28
P P o o s s s s i i b b l l e e S S o o l l u u t t i i o o n n
•
See the solution for Picture is too dark.
•
See the solution for Picture is too bright.
•
Perform the White Balance test with the light source
connected to the scope. Use metal-halide, xenon, or
LED lighting (no fluorescent lighting).
•
Increase the camera Light Level with the camera
head.
•
Increase the light source output.
•
Check the fiberoptic light cable for excessive broken
fibers.
•
Decrease the camera Light Level.
•
Decrease the light source output.
•
Turn off the DRE Desaturation function.
•
Plug the electrocautery generator into a separate
electrical outlet and separate the 1588 AIM console
power cord from the electrocautery power cord.
•
Separate the camera cable from the electrocautery
cable.
•
Reposition the electrocautery grounding pad on the
patient.
•
Confirm all cable connectors are securely attached.
•
Check for and replace faulty video cables.
•
Check to ensure that all devices in the video system
are plugged in and powered on.
•
Check the connector on the camera-head cable for
broken pins.
•
Detach the camera head from the console and
reconnect.
•
Turn off the console, wait 3 seconds, and turn it back
on.
•
Release the scope from the coupler and then
reconnect it. Make sure the scope is seated correctly
in the coupler.
P29924F