Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 94
Resistance to interference
• Note on device type DX (only devices with a DF-1/IS-1 connection): The EMC
requirements are met as long as atrial sensitivity is set to 1.0 mV (factory settings)
or values ≥ 1.0 mV. Measures must be taken to assure interference-free therapy if
more sensitive values are set.
• Note on device type HF: In the case of unipolar sensing, the requirement for inter-
ference voltages of ≤ 0.3 mV (tip to tip) is met.
Common mode rejection ratio
Rate
Common mode rejection ratio
Atrium: DX*
Atrium: DR,
HF
16.6 Hz
58 dB
53 dB
50 Hz
55 dB
55 dB
60 Hz
56 dB
56 dB
* only devices with a DF-1/IS-1 connection.
ATP amplitude
A burst was measured at 500 Ω, an amplitude of 7.5 V (tolerance ±1.5 V), pulse width of
1.5 ms, R-S1 interval of 300 ms and an S1 count of 5:
ATP amplitude
Measured
minimum
RV
7.67 V
LV
7.67 V
Automatic sensitivity setting
Measurement of actual values and test signal wave shape: standard triangle. For the
device type DX, the programmed atrial sensitivity is intensified by a factor of 4.
Sensitivity
Value
A: positive
0.2 mV
A: negative
DX: A: positive
0.2 mV
DX: A: negative
en • English
V right: VR, DR,
V left: HF
HF
64 dB
66 dB
64 dB
66 dB
64 dB
68 dB
Measured
Mean value
maximum
7.67 V
5.00 V
7.67 V
4.99 V
Tolerance
Measured value
0.2 ... 0.5
0.24 mV
0.24 mV
0.2 ... 0.52
0.05 mV
(0.05 to 0.13)
0.05 mV
Sensitivity
Value
RV: positive
0.5 mV
RV: negative
LV: positive
0.5 mV
LV: negative
Shock energy / peak voltage
With shock path: RV to housing + SVC
Shock energy 
Tolerance peak
(Tolerance)
voltage
1 J (nn ... nn)
90 ... 120 V
20 J (nn ... nn)
410 ... 450 V
45 J (nn ... nn)
620 ... 690 V

Battery Data

Battery characteristics
The following data is provided by the manufacturers:
Manufacturer
Battery type
Battery ID number shown on the
programmer
Device type
Battery voltage at ERI
Charge time at BOS
Charge time at ERI
Usable capacity until ERI
Usable capacity until EOS
21
Tolerance
Measured value
0.3 ... 0.7
0.48 mV
0.40 mV
0.3 ... 0.7
0.48 mV
0.56 mV
Measured value
Measured value
Shock energy
Peak voltage
0.82 J
94 V
18.1 J
444 V
40.4 J
663 V
GREATBATCH, INC.
LITRONIK GmbH & Co
Clarence, NY 14031
01796 Pirna, Germany
GB 2992
LiS 3411 RA
3
5
VR, (DX), DR, HF
2.5 V
2.85 V
10 s
10 s
12 s
12 s
1590 mAh
1590 mAh
1730 mAh
1730 mAh