2.7.1
Replacing fuses
DANGER
Danger to life through electric shock!
Unplug the device before changing the fuses.
Specified fuses: T2.00 AH/250 V~
Use a small screwdriver to release the clip on the fuse holder.
Pull out the fuse holder.
Replace both fuse sets.
Reinstall the fuse holder so that it audibly clicks into place.
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical service.
Note
Follow the country-specific regulations of the fuse.
Repairs
For repairs (except changing the fuse), contact the manufacturer or an
authorized repair center. Authorized repair centers can be inquired about
from the manufacturer.
For a fast processing of your service requests, send in the product indicat-
ing:
Article number (REF)
Serial number (SN)
Detailed description of defects
Note
If a component in the visualization system is replaced (e.g. service or
upgrade), the system commissioning must be performed again. It is also
recommended to perform servicing after any maintenance or replacement
of a device.
3.
Reprocessing procedure
3.1
General safety notes
Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for sterile pro-
cessing.
Note
Successful processing of this medical device can only be ensured if the pro-
cessing method is first validated. The operator/sterile processing technician
is responsible for this.
Note
For up-to-date information about reprocessing and material compatibility,
see also B. Braun eIFU at eifu.bbraun.com
3.2
Reusable products
Influences of the reprocessing which lead to damage to the product are
not known.
A careful visual and functional inspection before the next use is the best
opportunity to recognize a product that is no longer functional, see
Inspection.
3.3
Cleaning/disinfection
3.3.1
Product-specific safety instructions for the reprocessing pro-
cedure
Risk of electric shock and fire hazard!
Unplug the device before cleaning.
Do not use flammable or explosive cleaning or disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to the product due to improper reprocessing!
Use only wipe disinfection for cleaning and disinfecting the camera
control unit.
Do not, under any circumstances, clean or disinfect the product in an
ultrasonic cleaning bath.
Use cleaning and disinfecting agents which are suitable and approved
for the present product.
Observe the manufacturer's cleaning and disinfecting instructions
regarding concentration, temperature and exposure time.
Do not sterilize the product under any circumstances.
Risk of infection for patient and/or user!
Ensure that there is no residue from cleaning and disinfecting agents
on the product.
Avoid insufficient or incorrect cleaning and disinfection of the product
and accessory.
Risk of defects, consequential damage or a shortened product service life!
Follow and observe the manufacturer's requirements for reprocessing.
Risk of electric shock and damage to the device from moisture!
Carry out the wipe disinfection with damp cloths only. Never use drip-
ping wet cloths.
Remove excess cleaning solution from the device immediately.
Only reconnect the camera control unit to power when all cleaned
parts are completely dry.
Note
The housing of the Full HD CMOS camera control unit is only suitable to be
wiped off with a moist cloth (wipe disinfection). Do not immerse into liquid.
Note
Do not sterilize the camera control unit or reprocess it in a cleaning/disin-
fecting machine.