Guidance and manufacturer's declaration – electromagnetic immunity
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations on
for 0,5 cycle
power supply input lines
IEC 61000-4-11
40 % U
(60 % dip in U
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 s
Power frequency
3 A/m
(50/60 Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
<5 % U
T
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
T
)
(>95 % dip in U
T
for 5 s
30 A/m
Vercise Gevia™ Information for Prescribers
Electromagnetic Compatibility
Mains power quality
should be that of a typical
)
T
commercial or hospital
environment. If the user
of the Vercise Gevia DBS
System requires continued
operation during power
)
mains interruptions, it
T
is recommended that
the Vercise Gevia DBS
System be powered from
an uninterruptible power
supply or a battery.
)
T
)
T
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Magnetic fields from
common appliances are
not expected to affect the
device.
91168753-02 Rev A 13 of 505