Vercise Gevia™ Information for Prescribers
Guidance and manufacturer's declaration – electromagnetic immunity
The Vercise Gevia DBS System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Vercise Gevia DBS System should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Vercise Gevia™ Information for Prescribers
91168753-02 Rev A 14 of 505
Compliance Level
Electromagnetic Environment –
guidance
Portable and mobile RF communications
equipment should be used no closer
to any part of the Vercise Gevia DBS
System, including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
3 Vrms
d=1.2√P
150 kHz to 80 MHz
Outside ISM and
amateur radio
bands
a
6 Vrms
d=0.58√P
150 kHz to 80 MHz
Inside ISM and
amateur radio
bands
a