Alarm Messages, continued
Severity
Alarm Header
Impella Flow Reduced
Impella Position Unknown
(Impella 2.5, Impella CP, Impella
CP with SmartAssist, and Impella
5.5 with SmartAssist)
Impella Position Unknown
( Impella 2.5, Impella 5.0, Impella
LD, Impella CP, Impella CP with
SmartAssist, Impella 5.5 with
SmartAssist)
Placement Monitoring
Disabled
Placement Signal Lumen
Blocked
( Impella 2.5 and Impella CP)
Preventing Retrograde Flow
Purge Cassette Incompatible Contact Abiomed Service to update Impella
Purge Flow Decreased
Purge Flow Increased
Purge Volume Low
Surgical Mode Enabled
Transfer to Standard
Configuration
( Impella 2.5 and Impella CP)
Optical Sensor Not
Supported (Impella CP with
SmartAssist and Impella 5.5 with
SmartAssist)
Unexpected Controller
Shutdown
Update Cardiac Output
Automated Impella Controller with SmartAssist
Action
1. Reduce P-Level
2. Check Impella position.
3. Check filling and volume status.
Impella position unknown due to low pulsatility,
assess cardiac function.
Confirm Impella position with imaging.
Monitor Impella position with imaging or re-enable
Placement Monitoring under MENU and Settings/
Service.
Placement and Suction Monitoring are suspended.
Aspirate with syringe, then flush.
Impella P-level has increased to prevent retrograde
flow.
1. Consider increasing target P-level.
2. For weaning, disable Retrograde Flow Control
through MENU soft key.
Controller.
The purge flow has decreased by
2.5 mL/hr or more.
This is a notification only; no action is required.
The purge flow has increased by
2.5 mL/hr or more.
This is a notification only; no action is required.
1. Press the PURGE MENU soft key then select
Change Purge Fluid Bag.
2. Follow the instructions to change the purge fluid.
Impella pump stopped. Purge system running.
'Impella Stopped' alarm disabled.
To exit this mode start Impella pump.
Press the PURGE MENU soft key, then select Transfer
to Standard Configuration.
Placement and Suction Monitoring are Not Available.
1. Monitor Impella position with imaging.
2. Switch to Controller which supports optical
pressure sensor.
Switch to backup controller if condition persists.
Recommend entering a new reference Cardiac
Output. Press MENU soft key then select Enter
Cardiac Output.
®
Cause
Motor speed has been reduced in response to suction.
Impella Catheter position unknown due to low
pulsatility.
Impella Catheter position unknown detected by
algorithm.
User has disabled Placement Monitoring.
Placement signal pulsatility is low and placement signal
amplitude is high (>130mmHg) for 30 consecutive
seconds.
Retrograde flow has been detected and minimum
motor speed has been increased to more than target
P-level.
Incompatible purge cassette RFID version.
Purge flow has decreased by ≥2.5 mL/hr.
Purge flow has increased by ≥2.5 mL/hr.
There are 30 mL (in addition to 5% of the starting bag
volume) or fewer remaining in the purge fluid bag.
Surgical Mode has been enabled to silence "Impella
Stopped" alarm at P-0.
Follow instructions or press MUTE ALARM to clear the
alarm for 30 minutes.
The optical sensor is not supported with the current
version of the controller hardware.
The pump will still operate without placement or
suction monitoring.
Unexpected restart of controller due to software of
hardware failures.
Impella has detected a significant change in vascular
state.
15