Operation failure hazard warning
If an error message appears while a sterilisation cycle is
running, it will be necessary to repeat the cycle.
Accidentally dropping the device may deform it and
compromise proper operation. If this happens, have a full
check done by technical support.
If there are instruments that are joined together, they
must be divided or located in the most ventilated and
spacious place possible.
Lab coats or other reusable fabrics must be washed and
dried after use and before sterilisation to remove any
organic material and increase the "life" of the fabric by
restoring its natural water content (i.e. humidity level).
Comply with the load limits defined in this manual for
EN
each type of sterilisation load.
Crushing of body parts hazard warnings
The equipment has motorised door closure. Only hold
the door by the handle, or push it from the outside.
Be careful of your finger/hand position while closing the
door. Do not leave fingers or other body parts between
the door and the machine while it is closing.
Burn hazard warnings
To prevent possible burns, extract the trays with the de-
dicated pincers provided.
Before cleaning any surface, make sure that it has cooled
down.
Infect hazard warnings
The autoclave performs a sterilisation process. However,
the user must guarantee the effectiveness of the
sterilisation process by following the guidelines as well as
the data provided by the autoclave.
Therefore, strictly follow the protocols defined in this
manual to ensure that the process continues to be
effective.
Warnings against the risk of operator infection
Despite the water in the drain tank undergoing the
sterilization process, there is still a risk of biological
contamination. It is therefore prohibited to touch the drain
tank water. For disinfection follow the recommended
procedures in Section 5.
Note
The device, when on, must always have the tray holder
and tray on the lower shelf.
44
1.4 Operating and transport environmental conditions
I
n its packaging, the equipment can withstand the following
environmental conditions for no more that 15 weeks:
Ambient temperature from -20°C to +70°C
Relative humidity from 10% to 90%
Atmospheric pressure from 500 to 1060 mBar.
The appliance must be used in the following environmental
conditions:
- internal use
- altitude up to 3000 m
- Atmospheric pressure from 700 hPa to 1060 hPa
- temperature from 5°C to 40°C
- max relative humidity 85%
- max mains voltage variation ± 10%
- installation category (overvoltage category) II
- pollution degree rating 2
- ambient lighting 500 lx
1.5 Environmental disposal instructions
Pursuant to directives 2012/19 EC and 2011/65 EC on waste
electrical and electronic equipment and the restriction of the
use of certain hazardous substances in electrical and electronic
equipment, waste cannot be disposed of as urban waste, but must
be collected separately.
When purchasing new equipment of an equivalent type on a one-
for-one basis, the equipment that has reached the end of its life
must be returned to the dealer for disposal. Regarding reuse,
recycling or the other forms of waste recovery mentioned above,
the manufacturer must carry out the steps defined by the individual
national legislation.
Adequate separate collection for future use of the equipment
assigned for recycling, treatment and environmentally friendly
disposal helps to prevent possible negative effects on the
environment and on human health, and promotes recycling and/or
reuse of the materials that make up the equipment.
The crossed-out bin symbol (see symbols on page 38) on the device
or on its packaging indicates that the product must be disposed of
separately from other waste at the end of its life.