6.6 Message solutions
Code
phase
Msg001
Standby or cycle start
Msg002
Standby or cycle start
Msg003
Pin entry
Msg004
Water loading
Cycle start
Msg005
Vacuum Test start
Msg008
Machine switch-on
Msg009
EN
Machine switch-on
Msg010
Machine switch-on
Msg011
Msg012
Machine switch-on
Machine switch-on
Msg013
Machine switch-on or
Msg014
cycle start
Pin entry
Msg016
Msg017
Pin entry
Machine switch-on
Msg018
7
Techical specifications
Power supply voltage
Mains frequency
Power
Maximum current consumption
Fuse (6.3x32)
Noise level
Electric shock protection class
IP degree of protection
Operationg temperature
Chamber volume
Usable chamber volume
Feed tank volume
Weight per unit area of empty machine
Weight per unit area at full load
Empty machine weight
Fully loaded machine weight
Safety valve calibration pressure
7.1 Water specifications (distilled or demineralized)
Conductivity at 20°C
Chlorides
pH at 20°C
8
Spare parts codes
6046100101 : Tray handle
6048100023 : 17 L tray
6048100024 : 22 L tray
6048100019 : Bio-X-Filter
6048100018 : Door seal
6048100015*: Maintenance kit 1000 cycles
6048100138*: Maintenance kit 5000 cycles 17 liters
6048100139*: Maintenance kit 5000 cycles 22 liters
* The annual maintenance must only be carried out by an authorised and qualified service technician
76
Message
Feed tank empty
Feed tank ful
Incorrect Pin
Poor water quality
Door open
Chamber temperature
over limit
Serial number inconsistency
between CPU and MB
Lubrificate the automated
system of the door
Replace Bio-X filter
Conduct maintenance
Incorrect date
Do vacuum tes t
Pin too short
Duplicate PIN
MB FW revision incompatible
with CPU FW
Hygoclave 50
220 V
230 V
60 HZ
50/60 HZ
1880 W
2050 W
8,5 A
8,9 A
T12 A H 250v
approx. 17 litres
approx. 11 litres
202 kN/m2
237 kN/m2
57 kg
67 kg
2,5 bar
0÷15 green µS/cm
<2.0 mg/l
5.0 - 7.0
Possible solution
Fill the feed tank
Empty the discharge tank
Enter the correct Pin
Change the water in the feed tank
with better quality water
Close the door
Repeat Vacuum Test when the chamber temperature
returns below the allowed limit (45°C)
Contact The Dürr Dental Technical Service
Lubrificate the autoamted system according to the instructions
in par. 5.1.9 of the user manual
Replace the Bio-X filter with original Dürr Dental spare part
Contact service technician for routine maintenance
Set clock/calendar of the machine again. If the problem persists,
contact the Dürr Dental Technical Service
Do vacuum tes t
Entry 4-digit PIN
Entry correct PIN
Contact the Dürr Dental Technical Service
240 V
220 V
50 HZ
60 HZ
2230 W
2010 W
9,3 A
9,1 A
<60dB (A)
Class I
IPX0
From 5°C to 40°C
5,2 litres
15÷40 yellow µS/cm
> 40 red µS/cm
Hygoclave 50 Plus
230 V
240 V
50/60 HZ
50 HZ
2200 W
2400 W
9,6 A
10,0 A
T12 A H 250v
approx. 22 litres
approx. 14 litres
216 kN/m2
251 kN/m2
61 kg
71 kg
2,5 bar