6.4 Reset procedures
Reset procedure "A"
If the cycle is "running", reset the error by selecting the confirm icon on the safety screen and then select the Home icon to return to the initial
screen.
Reset procedure "B"
Reset the error be selecting the confirm icon on the safety screen.
N.B. In the event of faults that compromise sterilisation/test cycle execution, the cycle selection icons on the Home screen will be
disabled.
6.5 Troubleshooting
Code
There is a malfunction in the motorised door
ER 001
closing/opening system.
There is a malfunction in the motorised door
ER 002
closing/opening system.
There is a malfunction in the motorised door
ER 003
closing/opening system.
There is a malfunction in the motorised door
ER 004
closing/opening system.
The water quality is very poor.
ER 006
ER 007
Floating feed tank in unexpected condition.
There is a leak in the machine plumbing circuit
ER 100
or a malfunction in the vacuum pump.
ER 101
There is a leak in the machine plumbing circuit.
There is a leak in the machine plumbing circuit
ER 110
or a malfunction in the vacuum pump.
There is a leak in the machine plumbing circuit
ER 120
or a malfunction in the water injection pump.
There is a leak in the machine plumbing circuit
ER 121
or a malfunction in the water injection pump.
ER 130
There is a malfunction in the machine discharge system.
There is a leak in the machine plumbing circuit
ER 140
or a malfunction in the water injection pump.
There is a leak in the machine plumbing circuit
ER 141
or a malfunction in the water injection pump.
There is a leak in the machine discharge system or a
ER 150
chamber loading error.
There is a malfunction in the sterilisation
ER 151
condition holding system or a chamber
loading error.
Possible cause
Possible solution
Reset the machine.
Repeat the door closing procedure.
If the problem persists, contact technical support.
Reset the machine.
Very poor water quality: empty the feed tank and refill it with better
quality water.
If the problem persists, contact technical support.
Reset the machine. Check the condition of the floating feed tank.
If the problem persists, contact Technical Support.
Reset the machine. Check that the door seal is clean and positioned
correctly, then repeat the selected program. If the problem persists,
contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and repeat the selected
program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) the tray support position and the tray position in the chamber;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.