Code
There is a malfunction in the sterilisation condition
ER 152
holding/control system.
There is a malfunction in the sterilisation condition
ER 153
holding/control system.
There is a malfunction in the sterilisation condition
ER 154
holding/control system.
There is a malfunction in the machine discharge
ER 160
system.
EN
There is a leak in the machine plumbing circuit
ER 171
or a malfunction in the vacuum pump.
There is a malfunction in the machine
ER 172
drying system.
There is a leak in the machine plumbing circuit
ER 173
or a malfunction in the vacuum pump.
There is a malfunction in the pressure balancing
ER 180
system.
There is a leak in the machine plumbing circuit or a
ER 200
malfunction in the vacuum pump.
ER 201
There is a leak in the machine plumbing circuit.
ER 202
There is a leak in the machine plumbing circuit.
A Vacuum Test is running in chamber conditions
ER 203
other than ambient temperature.
There is a malfunction in the pressure balancing
ER 204
system.
Leak in the hydraulic circuit of the machine or
ER 310
malfunctioning of the vacuum pump.
Leak in the hydraulic circuit of the machine or
ER 340
malfunctioning of the water injection pump.
Leak in the hydraulic circuit of the machine or
ER 341
malfunctioning of the water injection pump.
ER 342
The safety valve has opened during test execution.
ER 355
The test execution time has been exceeded.
ER 360
Malfunction in the machine discharge system.
Malfunction in the machine drying system.
ER 370
Malfunction in the machine pressure balancing
ER 380
system.
AL 001
The reset button was pressed.
AL 002
A blackout has occurred.
74
Possible cause
Possible solution
Reset the machine. Check:
a) that there is water in the feed tank;
b) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) the tray support position and the tray positions in the chamber;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check the amount of load in the chamber and
then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and
the Bio-X filter condition, then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and
the Bio-X filter condition, then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check that the door seal is clean and positioned correctly, the repeat
the selected program.
If the problem persists, contact technical support.
Reset the machine.
Repeat the Vacuum Tes, making sure that the chamber is at ambient
temperature. If the problem persists, contact technical support.
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine. Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Replace the safety valve.
Check the condition of the safety valve and replace it if necessary.
Reset the machine. Check the amount of load in the chamber and
then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and the Bio-X filter
condition. Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Reset the machine. Check:
a) that the power cable is connected correctly;
b) that the is mains power in the system.
Then repeat the selected program.
If the problem persists, contact technical support.