Code
There is a malfunction in the feed tank water
AL 003
filling system.
AL 100
There is a malfunction in the chamber probe reading circuit.
AL 101
There is a malfunction in the generator probe reading circuit.
AL 102
There is a malfunction in the pipe bundle probe reading circuit.
There is a malfunction in the pressure transducer
AL 103
reading circuit.
There is a malfunction in the door closing/opening
AL 104
control system.
There is a malfunction in the door closing/opening
AL 105
control system.
There is a malfunction in the temperature
AL 200
control system.
There is a malfunction in the temperature
AL 201
control system.
There is a malfunction in the temperature
AL 202
control system.
There is a malfunction in the pressure
AL 203
control system.
There is a malfunction in the electronic system.
AL 300
AL 301
There is a malfunction in the electronic system.
There is a malfunction in the electronic system.
AL 302
AL 303
There is a malfunction in the electronic system.
There is a malfunction in the electronic system
AL 304
AL 305
There is a malfunction in the electronic system.
AL 306
There is a malfunction in the electronic system.
AL 307
There is a malfunction in the electronic system.
There is a malfunction in the electronic system.
AL 308
There is a malfunction in the electronic system.
AL 309
There is a malfunction in the electronic system.
AL 310
AL 311
There is a malfunction in the electronic system.
AL 401
Generator malfunction.
Band malfunction.
AL 402
Possible cause
Possible solution
Reset the machine. Check that the water filling hose is positioned and
connected correctly, then repeat the selected procedure.
If the problem persists, contact technical support.
If water comes out of the overflow fitting at the rear,
contact technical support.
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Contact technical support.
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
EN
75