product specific details. These labels will facilitate reordering of the device and its safe use within
a MR environment. The labels can be attached to the patient record.
Fold out pages: The image 1 represents the most complex tracheostomy tube
model.
1
tube
2
adjustable neck flange
2a
lever
2b
push button with spring element
2c
wings with eyelets
3
HVLP cuff
inflation line
3a
3b
pilot balloon with check valve
3c
female standard Luer connector
scale for axial orientation
4
5
standard 15mm connector
6
suction line
suction line port with female standard Luer
6a
connector
(1) Tracheostomy Tube:
All tubes are curved and feature a smooth, round tip at the distal end (inside the patient).
The tracheostomy tube (1) has a scale for axial orientation (4) behind the 15 mm connector
(5) indicating the position of the tube relative to the adjustable neck flange (2). This scale has no
measuring function.
REF 450-P, REF 451-P: The tube has a radiopaque metal spiral-reinforcement.
REF 460-P, REF 461-P, REF 470-P, REF 471-P: The tube is clear and has a radiopaque
stripe.
TRACOE vario extract tracheostomy tube with suction line (REF 470-P, and REF 471-P)
allows suctioning of secretions from the subglottic space.
The standardized 15 mm connector (5) is permanently attached to the tube and is intended
for connecting the tracheostomy tube to external devices with a female standardized 15 mm
connectors e.g. connection to mechanical ventilation, HME.
(2) Adjustable Neck Flange:
With the adjustable neck flange (2) the insertion depth of the tracheostomy tube can be
adapted to the patient's anatomical requirements.
The flange includes two flexible wings with eyelets (2c) for attachment of the neck strap
(12). The wings can be rotated within a range of approximately 60° to 180°. This permits universal
adjustment to meet the patient's requirements.
The product code (REF), clinical size (size), internal diameter (ID), external diameter (OD),
length range (TL) of the tube and MR Safety symbol are all indicated on the neck flange.
9
inserter
9a
handle with screwing function
10
guiding catheter
10a
silicone sleeve
11
obturator
12
neck strap
lubricant gel
13
14
adaptors
EN