(3) High-Volume-Low-Pressure (HVLP) Cuff:
The HVLP-cuff (3) is located on the distal end of the tracheostomy tube and directly con-
nected to the inflation line (3a).
The proximal end of the inflation line includes a pilot balloon (3b), with incorporated
EN
self-sealing check valve and a female Luer connector (3c).
The HVLP-cuff is inflated with air only.
The pilot balloon (3b) displays the cuff resting diameter (CD) and size, where appropriate.
(6 – 6a) Subglottic Suction Line:
TRACOE vario extract tracheostomy tubes (REF 470-P, REF 471-P) include a subglottic
suction line (6) that is integrated into the wall of tracheostomy tube and exits through an aperture
at the lowest possible position above the cuff.
The proximal end of the suction line includes a standard female Luer connector (6a) port for
connection to an external accessory device used in subglottic suctioning or for the air/oxygen
supply for ACV. For subglottic suctioning, additional adaptors (14) can be used for connection.
The subglottic suctioning port (6a) can be closed by using the attached cap.
(9 –10 a) Minimally Traumatic Insertion System:
The combination of the inserter (9) and guiding catheter (10) with silicone sleeve (10a)
constitute the minimally traumatic insertion system.
The silicone sleeve bridges the gap between the conical end of the inserter and the distal
end of the tube.
The minimally traumatic insertion system is used for the placement of the vario tracheos-
tomy tube using the Seldinger technique.
(11) Obturator:
The non-perforated obturator (11) has a smooth, round, conical tip at the distal end. The
obturator is used for re-insertion of the tracheostomy tube for a tracheostoma that does not
require the Seldinger technique.
(12) Neck Strap:
The neck strap (12) is a soft strip of padded fabric that wraps around the patient's neck.
The ends of the strap include hook-and-loop fasteners that are inserted through the eyelets
of the neck flange to secure the tracheostomy tube in position.
The frequency of change is determined by the physician or healthcare professional.
(13) Lubricating Gel:
The lubricating gel (13) can be used for the insertion of the minimally traumatic insertion
system or obturator.
Supplementary Products:
Products, which can be used in combination with the vario tracheostomy tube are listed in
section "Supplementary Products".
3.
MRI Safety Information
MR Unsafe REF 450-P and REF 451-P
Keep the vario tracheostomy tube REF 450-P and REF 451-P outside the MRI scanner room.
MR Conditional REF 460-P, REF 470-P, REF 461-P and REF 471-P